FDA recall Z-1561-2022

Synapse Biomedical Inc · Class I · device

Product

NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.

Reason for recall

Due to a defect identified on the EPG controller printed circuit board assembly, device performance may degrade over time. Depending on the failure mode, potential adverse events include pain, muscle fatigue and decreased breathing effectiveness, arrhythmia or cardiac arrest, or the device may stop unexpectedly.

Distribution

distribution to US states of AL, CO, IL, FL, GA, MD, MN, NC, NJ, OH, PA, and SC; Germany and Kuwait. ***Updated 3/28/23 - Product was not distributed to United Arab Emirates. Product was distributed to Kuwait.*** Updated 9.14.23 - 1 patient in Germany, 1 patient in the Czech Republic, 1 distributor in Kuwait

Key facts

Status
Terminated
Initiation date
2022-07-11
Report date
2022-08-31
Termination date
2023-10-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Oberlin, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1561-2022