Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351
Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.
Domestic: IN, MA, CA, NJ, TX, NE, NY, VA, MD, MN, OR, IL, CT, DC, MS. Foreign: Albania, Argentina, Australia, Bangladesh, Bolivia, Bosnia & Herzegovina, Cameroon, Canada, Colombia, Cyprus, Denmark, Egypt, Estonia, Fiji, France, Germany, Greece, Hong Kong, Iraq, Italy, Ireland, Jordan, Kazakhstan, Kenya, Lithuania, Luxembourg, Mauritius, Mexico, Myanmar, Nepal, New Zealand, Pakistan, Palestine, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, The Netherlands, Turkey, UK, Ukraine, USA, and Vietnam.