FDA recall Z-1572-2026

Howmedica Osteonics Corp. · Class II · device

Product

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Reason for recall

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Distribution

Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;

Key facts

Status
Ongoing
Initiation date
2026-02-02
Report date
2026-03-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mahwah, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1572-2026