# FDA recall Z-1577-2025

> **Maquet Cardiovascular, LLC** · Class II · device recall initiated 2025-03-10.

## Product

Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).

## Reason for recall

Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.

## Distribution

Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.

## Key facts

- **Recall number:** Z-1577-2025
- **Recalling firm:** Maquet Cardiovascular, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-03-10
- **Report date:** 2025-04-23

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Wayne, NJ, United States