FDA recall Z-1579-2023

CareFusion 303, Inc. · Class II · device

Product

BD Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES (Med ES Main Tower), REF: 352; BD Pyxis Anesthesia Station ES, REF: 327; BD Pyxis CII Safe ES Tower Main, REF: 1116-00; BD Pyxis CII Safe ES, Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01

Reason for recall

Automated dispensing cabinet software is experiencing: 1) ES device download failure resulting in partial patient information loss that leads to 2) partial loss of patient transaction data, seen as a server upload processing error. Outdated/incorrect information display may contribute to the removal of medication to which the patient is allergic, incorrect dose amount, or discontinued medication.

Distribution

US Nationwide distribution in the states of KS, NY, NC, LA, MA, MI.

Key facts

Status
Ongoing
Initiation date
2023-04-05
Report date
2023-05-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States