# FDA recall Z-1587-2026

> **Datascope Corp.** · Class III · device recall initiated 2026-02-06.

## Product

Battery Charging Station;  Model: 0998-00-0802;

## Reason for recall

The Battery Charging Station is an optional dual-bay  charger used to charge the Cardiosave IABP Lithium-  Ion batteries when they are not being used to  operate the IABP. Getinge has identified that a  protruding screw in the left battery bay limits full  insertion of certain batteries and prevents proper  charging. The right bay of the Battery Charging  Station is not affected by this issue.

## Distribution

Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.

## Key facts

- **Recall number:** Z-1587-2026
- **Recalling firm:** Datascope Corp.
- **Classification:** Class III
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-02-06
- **Report date:** 2026-03-25

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Mahwah, NJ, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1587-2026

## Citation

> AI Analytics. FDA recall Z-1587-2026. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-1587-2026. Source: US FDA. Licensed CC0.

---