FDA recall Z-1592-2024

ACTIM OY · Class II · device

Product

Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane

Reason for recall

Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.

Distribution

US Nationwide distribution in the state of Connecticut.

Key facts

Status
Ongoing
Initiation date
2023-11-23
Report date
2024-04-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Espoo, Finland

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1592-2024