# FDA recall Z-1595-2021

> **Roche Diagnostics Operations, Inc.** · Class II · device recall initiated 2021-04-07.

## Product

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

## Reason for recall

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

## Distribution

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.

## Key facts

- **Recall number:** Z-1595-2021
- **Recalling firm:** Roche Diagnostics Operations, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2021-04-07
- **Report date:** 2021-05-26

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Indianapolis, IN, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1595-2021

## Citation

> AI Analytics. FDA recall Z-1595-2021. Retrieved 2026-06-26 from https://api.ai-analytics.org/recall/Z-1595-2021. Source: US FDA. Licensed CC0.

---