FDA recall Z-1595-2023

Steris Corporation · Class II · device

Product

Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062

Reason for recall

In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.

Distribution

Worldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: China, Lithuania, Saudi Arabia, Korea, Republic of, Canada, Taiwan, United Kingdom, Spain, Australia, New Zealand, Spain, Brazil, Bahrain, Switzerland, Great Britain, South Africa, Hong Kong, Qatar.

Key facts

Status
Ongoing
Initiation date
2023-04-10
Report date
2023-05-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mentor, OH, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1595-2023