I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 31404-44, 31404-46, 31404-48, 31404-50, 31404-55, 31404-60, 31404-65, 31404-70, 31404-75, 31404-80, 31457-30, 31457-36, 31457-40, 31457-46, 31457-50, 31457-55, 31457-60, 31652-100, 31652-105, 31652-110, 31652-115, 31652-120, 31652-25, 31652-30, 31652-35, 31652-40, 31652-45, 31652-55, 31652-60, 31652-65, 31652-70, 31652-75, 31652-80, 31652-85, 31652-90, 31652-95, 31731-100, 31731-105, 31731-110, 31731-115, 31731-120, 31731-50, 31731-60, 31731-65, 31731-70, 31731-75, 31731-80, 31731-85, 31731-90, 31731-95, 31732-60, 31734-100, 31734-105, 31734-110, 31734-115, 31734-120, 31734-45, 31734-50, 31734-60, 31734-65, 31734-70, 31734-75, 31734-80, 31734-85, 31734-90, 31734-95. 2. Cortical Screw, Cannulated; Article Number
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1599-2026