FDA recall Z-1602-2021

Medtronic Inc. · Class II · device

Product

Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

Reason for recall

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2021-02-26
Report date
2021-05-26
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northridge, CA, United States