FDA recall Z-1603-2018

Philips Medical Systems Nederlands · Class II · device

Product

MultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Reason for recall

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

Distribution

Worldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latvia, Lebanon, Luxembourg, Malaysia, Martinique Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Qatar, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.

Key facts

Status
Terminated
Initiation date
2018-03-26
Report date
2018-05-09
Termination date
2020-06-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Best, N/A, Netherlands

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1603-2018