Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: GS3 5PK 1L AUS/ GTIN: 20763000252056 GS3 5PK 1L AUS/ GTIN: 20763000414362 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Worldwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1611-2021