# FDA recall Z-1615-2024

> **XENEX Disinfection Services Inc.** · Class II · device recall initiated 2024-03-18.

## Product

LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.

## Reason for recall

Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.

## Distribution

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.

## Key facts

- **Recall number:** Z-1615-2024
- **Recalling firm:** XENEX Disinfection Services Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-03-18
- **Report date:** 2024-05-01

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Antonio, TX, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1615-2024

## Citation

> AI Analytics. FDA recall Z-1615-2024. Retrieved 2026-07-20 from https://api.ai-analytics.org/recall/Z-1615-2024. Source: US FDA. Licensed CC0.

---