FDA recall Z-1617-2020

Haemonetics Corporation · Class II · device

Product

TEG Manager software versions used in conjunction with TEG 5000 are 1.1.0, 3.0.0, 4.0.0, 4.1.0 and 4.1.1. Intended: For in vitro coagulation studies

Reason for recall

Software defect in TEG Manager impacts the displayed alert for out of range test results. Due to this defect, reference range values received from TEG 5000 and displayed on the TEG Manager test result screen are rounded to the nearest whole number and lead to TEG Manager displaying an out of range alert when the test result is actually in range, or vice versa.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-02-19
Report date
2020-04-08
Termination date
2021-06-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Boston, MA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1617-2020