FDA recall Z-1620-2020

Stryker Neurovascular · Class II · device

Product

Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-115, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459210(17)211129(10)00000-00 Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-125, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459227(17)211129(10)00000-00 Stryker Neurovascular, AXS Vecta 71 Aspiration Catheter, Ref. Catalog No. INC-11129-132, Use By 2021-11-29, Rx Only, Sterile/EO, Manufactured For: Stryker Neurovascular 47900 Bayside Parkway, Fremont, CA 94538, UDI: (01)07613327459234(17)211129(10)00000-00 Product Usage: The AXS Vecta Aspiration System, including the AXS Vecta Aspiration Catheter, Aspiration Tubing Set, and VC-701 Cliq Aspirator Pump, is indicated in the revascularization of patients with acute ischemic stro

Reason for recall

The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.

Distribution

US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC. No US Govt. No Foreign countries.

Key facts

Status
Terminated
Initiation date
2019-04-10
Report date
2020-04-08
Termination date
2021-02-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Fremont, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1620-2020