FDA recall Z-1620-2024

SEASPINE ORTHOPEDICS CORPORATION · Class II · device

Product

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Reason for recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution

US: CA, FL, CO, TX

Key facts

Status
Ongoing
Initiation date
2024-02-28
Report date
2024-05-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Carlsbad, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1620-2024