FDA recall Z-1632-2018

Michigan Instruments, Inc. · Class III · device

Product

Thumper Cardiopulmonary Resuscitator, Model 1007CCV, Catalog No. 15370 Product Usage: The device is used to perform Cardiopulmonary Resuscitation (CRP) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.

Reason for recall

A performance issue with the chest depth needle valve (CDNV) may arise where turning the knob will not deliver any compression depth to the piston during the first 5 revolutions of the knob.The last ¿ to ¿ revolution of the knob could deliver a full 8cm of compression.

Distribution

Distributed to China.

Key facts

Status
Terminated
Initiation date
2018-02-13
Report date
2018-05-16
Termination date
2019-04-03
Voluntary/Mandated
Voluntary: Firm initiated
Location
Grand Rapids, MI, United States