FDA recall Z-1636-2024

Quidel Corporation · Class II · device

Product

QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20-U, 20 Tests

Reason for recall

There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.

Distribution

US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MT NC ND NE NH NJ NV NY OH OK OR PA SC SD TN TX UT VA WA WI WV OUS: United Arab Emirates, Belgium, Canada, Chile, China, Germany, Spain, United Kingdom, Ireland, Italy, South Korea, Lithuania, Qatar

Key facts

Status
Ongoing
Initiation date
2024-03-05
Report date
2024-05-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States