FDA recall Z-1645-2018

Deerfield Imaging, Inc. · Class II · device

Product

IMRIS IMRI 1.5T AND 3T S System Product Usage: The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designated to operative inside an RF shielded room to facilitate intraoperative and multi room use.

Reason for recall

The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction.

Distribution

Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada

Key facts

Status
Terminated
Initiation date
2018-04-04
Report date
2018-05-16
Termination date
2020-07-22
Voluntary/Mandated
Voluntary: Firm initiated