FDA recall Z-1650-2023

Philips Respironics, Inc. · Class I · device

Product

Trilogy Evo O2 and LifeVent EVO2, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, LA2100X15B, IA2100X15B, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, EU2100X15B, EU2100X19, EE2100X15B, and SP2100X26B

Reason for recall

Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

Distribution

US Nationwide. Global Distribution.

Key facts

Status
Ongoing
Initiation date
2023-05-01
Report date
2023-06-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Murrysville, PA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1650-2023