# FDA recall Z-1650-2026

> **Medline Industries, LP** · Class II · device recall initiated 2026-02-10.

## Product

Puracol and Puracol Plus Collagen Wound Dressings:    MSC8522	DRESSING,COLLAGEN,PURACOL,STRL,2" X 2"  MSC8622EP	DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"

## Reason for recall

Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.

## Distribution

Domestic distribution nationwide US. International distribution to Mexico and Panama.

## Key facts

- **Recall number:** Z-1650-2026
- **Recalling firm:** Medline Industries, LP
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2026-02-10
- **Report date:** 2026-04-01

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Northfield, IL, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1650-2026

## Citation

> AI Analytics. FDA recall Z-1650-2026. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-1650-2026. Source: US FDA. Licensed CC0.

---