FDA recall Z-1658-2023

Maquet Medical Systems USA · Class II · device

Product

QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-USA 701070397 BEQ-HMOD30000-USA 701050330

Reason for recall

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Distribution

Nationwide domestic distribution

Key facts

Status
Ongoing
Initiation date
2023-05-18
Report date
2023-06-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1658-2023