FDA recall Z-1666-2023

Medtronic Navigation, Inc. · Class I · device

Product

StealthStation Cranial Software, Model: 9735585, used with StealthStation S7/i7 systems

Reason for recall

In nonaxial/some axial exams if surgical plan utilized, Target Guidance selected, tip projection utilized, Navigate Projection enabled, software anomaly may occur during tumor resection/shunt placement/deep brain stimulation, where distance to/past target text no longer synchronized with navigation information, inaccurate value displayed; may result in prolonged/additional procedure, tissue injury

Distribution

US Distribution: PR. OUS (foreign) distribution: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, Colombia COLOMBIA, C¿TE D'IVOIRE, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, LATVIA, MEXICO, MOLDAVIA, NETHERLANDS, PAKISTAN, POLAND, PORTUGAL, REPUBLIC of KOREA, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES

Key facts

Status
Ongoing
Initiation date
2023-04-11
Report date
2023-06-14
Voluntary/Mandated
Voluntary: Firm initiated