FDA recall Z-1667-2023

NOXBOX LTD · Class I · device

Product

NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device

Reason for recall

Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.

Distribution

Worldwide distribution - US Nationwide and the countries of Bahrain, Bolivia, Brazil, Ecuador, Finland, France, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Panama, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Turkey, UAE, United Kingdom.

Key facts

Status
Terminated
Initiation date
2023-05-03
Report date
2023-06-21
Termination date
2024-10-31
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sittingbourne, N/A, United Kingdom

Primary source