FDA recall Z-1671-2020

Abbott Laboratories, Inc · Class II · device

Product

ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes.

Reason for recall

Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.

Distribution

Worldwide distribution  US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, ME, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, Puerto Rico and St. Croix US Virgin Islands. Countries of AF, AL, AO, AR, AM, AU, AT, AZ, BS, BD, BV, BE, BA, BW, BR, BN, BG, BF, CA, CL, CN, CO, CR, HR, CY, CZ, CD, DK, DO, EC, EG, SV, EE, ET, FI, FR, GE, DE, GR, GT, HN, HK, HU, IN, ID, IQ, IE, IL, IT, JM, JO, KZ, KE, KW, KG, LV, LB, LY, LT, LU, MK, MW, MY, MV, ML, MR, MX, MD, MA, MM, NA, NL, NZ, NG, NO, OM, PK, PA, PE, PH, PL, PT, PY, QA, RO, RU, RW, SA, SN, SG, SK, SI, ZA, ES, SE, CH, TW, TJ, TZ, TH, TT, TN, TC, TR, AE, UG, UK, UA, UY, VE, VN, YE, ZM, ZW, LC, MS, ME, VC, LK, RS, KY, CW, GZA, JRH, CI, KR.

Key facts

Status
Terminated
Initiation date
2019-05-01
Report date
2020-04-15
Termination date
2021-07-02
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irving, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1671-2020