Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c) REF 1-10013-111, SIZE 1-LEFT MED, 11MM; d) REF 1-10013-113, SIZE 1-LEFT MED, 13MM; e) REF 1-10013-159, SIZE 1-RIGHT MED, 9MM; f) REF 1-10013-160, SIZE 1-RIGHT MED, 10MM; g) REF 1-10013-161, SIZE 1-RIGHT MED, 11MM; h) REF 1-10013-163, SIZE 1-RIGHT MED, 13MM; i) REF 1-10013-209, SIZE 2-LEFT MED, 9MM; j) REF 1-10013-210, SIZE 2-LEFT MED, 10MM; k) REF 1-10013-211, SIZE 2-LEFT MED, 11MM; l) REF 1-10013-213, SIZE 2-LEFT MED, 13MM; m) REF 1-10013-259, SIZE 2-RIGHT MED, 9MM; n) REF 1-10013-260, SIZE 2-RIGHT MED, 10MM; o) REF 1-10013-261, SIZE 2-RIGHT MED, 11MM; p) REF 1-10013-263, SIZE 2-RIGHT MED, 13MM; q) REF 1-10013-309, SIZE 3-LEFT MED, 9MM; r) REF 1-10013-310, SIZE 3-LEFT MED, 10MM; s) REF 1-10013-311, SIZE 3-LEFT MED, 11MM; t) REF 1-10013-313, SIZE 3-LEFT MED, 13MM; u) REF
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
US Nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1673-2023