FDA recall Z-1674-2020

Beckman Coulter Inc. · Class II · device

Product

Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.

Reason for recall

Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.

Key facts

Status
Terminated
Initiation date
2019-12-13
Report date
2020-04-15
Termination date
2021-04-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States