FDA recall Z-1684-2022

Saranas, Inc. · Class II · device

Product

Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)

Reason for recall

Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.

Distribution

US Nationwide distribution including in the states of PA, MI, FL, WV, TX, NJ, GA, LA, NY, MA, CT, AZ, OH, TN, VA.

Key facts

Status
Ongoing
Initiation date
2022-08-10
Report date
2022-09-14
Voluntary/Mandated
Voluntary: Firm initiated
Location
Houston, TX, United States

Primary source