FDA recall Z-1688-2020

Abbott Laboratories · Class II · device

Product

"HEMOGLOBIN A1c CALIBRATORS *** REF 4P52-01 *** 307261/R03 *** S4P5X0 *** FOR USE WITH ARCHITECT" - Product Usage: use in the calibration of the Hemoglobin A1c assay.

Reason for recall

Calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. The manufacturer observed no impact to assay performance between the HbA1c concentration range of 5.33% to 6.87%. However, outside of this range, a bias of greater than 3% may be seen with patient results. The device failure that is associated with this recall can result in potentially negatively and positively biased patient test results for HbA1c. The firm is investigating the root cause.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, FKL GA, IA, ID, IL, IN, KS, LA, MN, MO, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, and WI; and countries of Austria, Germany, Italy, South Africa, and Spain.

Key facts

Status
Terminated
Initiation date
2019-05-22
Report date
2020-04-15
Termination date
2021-01-22
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irving, TX, United States