FDA recall Z-1724-2026

Medline Industries, LP · Class I · device

Product

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. MTS RH LFT HRT KIT, Medline SKU # 60010508; 2. MTS 3V W/GUIDE, Medline SKU # 60100701; 3. ANGIOGRAPHY KIT, Medline SKU # 60131209; 4. MTS LHK, Medline SKU # 60160075; 5. MTS LHK, Medline SKU # 60161954; 6. MTO LHK, Medline SKU # 60230148; 7. GENERAL SURGERY TRAY, Medline SKU # 00-500510I; 8. BIOBURDEN TEST KIT - 686, Medline SKU # BIOB686; 9. RADIO ANGIO UNIV HOSP PACK-LF, Medline SKU # DYNJ0376145M; 10. SPECIALS PACK-LF, Medline SKU # DYNJ0392706K; 11. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426O; 12. CARDIAC CATH PACK-LF, Medline SKU # DYNJ0451426P; 13. CARDIAC CATH PACK, Medline SKU # DYNJ07953N; 14. CATH PACK-LF, Medline SKU # DYNJ0803367J; 15. E.P. TRAY, Medline SKU # DYNJ19343J; 16. CARDIAC CATH PACK, Medline SKU # DYNJ20898T; 17. CATH LAB TRAY, Medline SKU # DYNJ24250G; 18. CATH LAB PACK-LF, Me

Reason for recall

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Key facts

Status
Ongoing
Initiation date
2026-02-27
Report date
2026-04-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1724-2026