# FDA recall Z-1729-2025

> **Remote Diagnostic Technologies Ltd.** · Class II · device recall initiated 2025-04-15.

## Product

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R,  00-1026-R, 00-1004, 00-1007

## Reason for recall

Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.

## Distribution

Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.

## Key facts

- **Recall number:** Z-1729-2025
- **Recalling firm:** Remote Diagnostic Technologies Ltd.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-04-15
- **Report date:** 2025-05-14

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Farnborough, N/A, United Kingdom