FDA recall Z-1746-2022

Qiagen Sciences LLC · Class II · device

Product

QIAcube Connect MDx, Model No. 9003070

Reason for recall

During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.

Distribution

Domestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2021-11-21
Report date
2022-09-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Germantown, MD, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1746-2022