FDA recall Z-1746-2025

Ethicon Endo-Surgery Inc · Class I · device

Product

Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35

Reason for recall

Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.

Distribution

US distribution nationwide. OUS distribution to EMEA, LATAM, APAC, and Canada.

Key facts

Status
Ongoing
Initiation date
2025-04-22
Report date
2025-05-28
Voluntary/Mandated
Voluntary: Firm initiated