# FDA recall Z-1751-2020

> **Beckman Coulter Inc.** · Class II · device recall initiated 2019-09-04.

## Product

Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators. The  assay provides a rapid screening procedure for presence of the analyte in urine.

## Reason for recall

Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of  certain compounds in urine may result in false positive or false negative patient test results.

## Distribution

Distributed throughout  the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.

## Key facts

- **Recall number:** Z-1751-2020
- **Recalling firm:** Beckman Coulter Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-09-04
- **Report date:** 2020-04-29
- **Termination date:** 2021-04-28

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Brea, CA, United States