# FDA recall Z-1751-2022

> **Alcor Scientific, Inc.** · Class II · device recall initiated 2022-07-20.

## Product

iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the  iSED Automated ESR Analyzer   PN: 112-13-002

## Reason for recall

Primary Pump Tubing Spare Part is incorrect.  If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube

## Distribution

US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea

## Key facts

- **Recall number:** Z-1751-2022
- **Recalling firm:** Alcor Scientific, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2022-07-20
- **Report date:** 2022-09-21

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Smithfield, RI, United States