FDA recall Z-1752-2020

Beckman Coulter Inc. · Class II · device

Product

Opiate 300 Ng Reagent (OP 1 X 250) for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems OP 300 Urine Calibrators, Catalogue No.475024. The OP assay provides a rapid screening procedure for determining the presence of OP and its metabolites in urine.

Reason for recall

Cross-reactivity testing may not meet the claims indicated in the Opiate 300 ng (OP)Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

Distribution

Distributed throughout the United States and to Puerto Rico. States include: CA, NC, OR, MI, GE, AL, SC, TX, NJ, IA, MS, IN, MA, NC, OH, AZ, LA, MA, NY, PA, KY, WV, MI, WA, CO, IL, VA, FL, KS, MD, ID, DE, NV, OK, AR, HI, WI, NY.

Key facts

Status
Terminated
Initiation date
2019-09-04
Report date
2020-04-29
Termination date
2021-04-28
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brea, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1752-2020