FDA recall Z-1755-2024

Johnson & Johnson Surgical Vision, Inc. · Class II · device

Product

Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.

Reason for recall

Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.

Distribution

U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.

Key facts

Status
Ongoing
Initiation date
2024-03-14
Report date
2024-05-08
Voluntary/Mandated
Voluntary: Firm initiated