FDA recall Z-1756-2018

Cook Inc. · Class II · device

Product

Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.

Reason for recall

Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.

Distribution

Distributed to accounts in 7 states: AL, CA, GA, NE, NV, NY, and TX. International distribution to Japan, Korea, and Taiwan.

Key facts

Status
Terminated
Initiation date
2018-03-29
Report date
2018-05-16
Termination date
2019-04-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Bloomington, IN, United States