FDA recall Z-1757-2024

Philips Ultrasound, Inc. · Class II · device

Product

5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.

Reason for recall

Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns

Distribution

Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates.

Key facts

Status
Ongoing
Initiation date
2024-03-23
Report date
2024-05-15
Voluntary/Mandated
Voluntary: Firm initiated