FDA recall Z-1761-2019

Teleflex Medical · Class I · device

Product

Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management

Reason for recall

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

Distribution

Worldwide - US Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2019-05-24
Report date
2019-06-26
Termination date
2022-02-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Morrisville, NC, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1761-2019