FDA recall Z-1762-2019

Teleflex Medical · Class I · device

Product

Hudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b) 5-22312, c) 5-22112, d) 5-22012, 2) Preformed 6.5 mm, Product Codes: a) 5-22313, b) 5-22213, c) 5-22013, d) 5-22113, 3) Preformed 7.0 mm, Product Codes: a) 5-22214, b) 5-22314, c) 5-22014, d) 5-22114, 4) Preformed 7.5 mm, Product Codes: a) 5-22215, b) 5-22315, 5) Preformed 8.0 mm, Product Codes: a) 5-22216, b) 5-22316, 6) Preformed 8.5 mm, Product Codes: 5-22217 Product Usage: Tracheal tube/airway management

Reason for recall

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

Distribution

Worldwide - US Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2019-05-24
Report date
2019-06-26
Termination date
2022-02-15
Voluntary/Mandated
Voluntary: Firm initiated
Location
Morrisville, NC, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1762-2019