FDA recall Z-1763-2025

ICU Medical, Inc. · Class II · device

Product

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software SUite v2.1 and v2.2

Reason for recall

Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

Distribution

Worldwide - US Nationwide distribution in the states of CA, FL, IA, KY, LA, MA, MS, NY, PA, TN, and VA. The country of Philippines.

Key facts

Status
Ongoing
Initiation date
2025-05-02
Report date
2025-05-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lake Forest, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1763-2025