FDA recall Z-1768-2026

Cardinal Health 200, LLC · Class II · device

Product

Cardinal Health Monoject Tuberculin Syringe, 1 mL Luer Lock, Catalog Number 1180100777

Reason for recall

The outer packaging and blister pack labeling identify the product as 1mL Luer Lock Tuberculin Syringe (Product Code 1180100777), however the syringes contained within the packages are U-100 Insulin syringes.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2026-02-27
Report date
2026-04-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukegan, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1768-2026