FDA recall Z-1776-2024

Medtronic Navigation, Inc. · Class I · device

Product

Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures.

Reason for recall

Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Republic Of Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, Nicaragua, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

Key facts

Status
Ongoing
Initiation date
2024-03-26
Report date
2024-05-15
Voluntary/Mandated
Voluntary: Firm initiated