# FDA recall Z-1777-2019

> **Maquet Cardiovascular Us Sales, Llc** · Class II · device recall initiated 2019-05-20.

## Product

Getinge 86-Series-Medical Washer-Disinfector   Part Number: S-86662003     Product Usage:  Usage:  The Getinge 86 Series Medical Washer-Disinfector is designed for use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.

## Reason for recall

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

## Distribution

Worldwide Distribution - US Nationwide GA, IL, IN, MI, NC, OR, PA, VA    Foreign:   Australia, Canada, China, France, Germany, India, Japan, Korea, Netherlands, Saudi Arabia, South Africa, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, Ukraine

## Key facts

- **Recall number:** Z-1777-2019
- **Recalling firm:** Maquet Cardiovascular Us Sales, Llc
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2019-05-20
- **Report date:** 2019-06-19
- **Termination date:** 2023-12-15

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Wayne, NJ, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1777-2019