FDA recall Z-1790-2020

CSL Behring GmbH · Class II · device

Product

Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997

Reason for recall

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

Distribution

IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan

Key facts

Status
Terminated
Initiation date
2020-03-02
Report date
2020-04-29
Termination date
2020-10-05
Voluntary/Mandated
Voluntary: Firm initiated
Location
Marburg, N/A, Germany

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1790-2020