# FDA recall Z-1790-2021

> **Terumo Cardiovascular Systems Corporation** · Class II · device recall initiated 2021-04-16.

## Product

The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. The TCM with options will also supply water for cardioplegia, freeze water for an ice supply, monitor temperatures in the patient and extracorporeal circuit, and allow gradient rewarming relative to a venous blood temperature.    Device Name / Model Number:   TCM II TUV, 115V (P/N 4415),   TCM II TUV, 220V (P/N 4416),     Catalog Number:   4416, 164940

## Reason for recall

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

## Distribution

Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam

## Key facts

- **Recall number:** Z-1790-2021
- **Recalling firm:** Terumo Cardiovascular Systems Corporation
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2021-04-16
- **Report date:** 2021-06-09
- **Termination date:** 2023-11-29

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Ann Arbor, MI, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1790-2021