# FDA recall Z-1793-2020

> **bioMerieux, Inc.** · Class II · device recall initiated 2020-01-24.

## Product

PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units),  P1028888-001 PSU (used in Zebra LP2824 Plus printers)     PSUs with the following date code; 0639XX through 1253XX (YYWWXX, where YY=Year, and WW=week).

## Reason for recall

The recall has been initiated due to potential for overheating or  a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located within the PREVI Isola instrument.

## Distribution

Worldwide distribution.  US nationwide including AU, BE, CH, DE, ES, FR, IT, HK/CN; MX, NL, RU, SE/FI, ZA.

## Key facts

- **Recall number:** Z-1793-2020
- **Recalling firm:** bioMerieux, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-01-24
- **Report date:** 2020-04-29
- **Termination date:** 2023-01-11

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Durham, NC, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1793-2020

## Citation

> AI Analytics. FDA recall Z-1793-2020. Retrieved 2026-06-24 from https://api.ai-analytics.org/recall/Z-1793-2020. Source: US FDA. Licensed CC0.

---