FDA recall Z-1800-2020

Bio-Rad Labs · Class II · device

Product

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage: The assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings.

Reason for recall

An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.

Distribution

US Nationwide distribution including the states of AZ, NY, LA, FL, CA, TX, IA, MS, UT, and IL.

Key facts

Status
Completed
Initiation date
2019-04-03
Report date
2020-05-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Benicia, CA, United States